I: Principles and role:
◆ UV resistant climate chamber using fluorescent UV lamp as light source, by simulating natural sunlight UV radiation and condensation of the material accelerated weathering test in order to obtain material weathering results.
◆ UV resistant climate chamber can simulate natural climate in the UV, rain, high temperature, high humidity, condensation, darkness and other environmental conditions, through to reproduce these conditions, combined into a loop, and it automatically cycles to complete . This is the UV aging chamber works.
First, the technical specifications
1, temperature range: RT +10 ℃ ~ 70 ℃
2, humidity range: ≥ 95% RH
3, the temperature fluctuation: ≤ ± 0.5 ℃
4, the temperature uniformity: ≤ ± 2 ℃
5, humidity fluctuation: ≤ ± 2%
6, humidity uniformity: ≤ ± 2%
7, the board temperature: 40 ℃ ~ 65 ℃
8, the distance between the tube: 35mm
9, sample and lamp distance: 50mm
10, lamp power: 40W
11, UV wavelength: 290nm ~ 400nm
12, analog Gel: Gel System time is adjustable
13, UV exposure, condensation simulation time range: 0 to 99 hours
14, support the model: 150 * 75 (mm)
15, model number: about 40
16, the test of time: 0 ~ 9999H adjustable
17, irradiance range: ≤ 50w/m2
Second, the structure
1, Box Case Material: Imported stainless steel SUS 304 # 1.2
2, the liner material: imported stainless steel plate SUS 304 # 1.2
3, the lid material: imported stainless steel plate SUS 304 # 1.2
4, in the studio were installed 8 on both sides of the UV lamp Philips
5, heating the inner tank heating, warming up quickly, the temperature distribution
6, two-way flip cover, ease of opening and closing
7, automatic water tank level
8, sample holder and liner done by the springs are made of aluminum alloy material.
9, the bottom chamber can be fixed with high quality PU wheel activity
10, the drainage system back to the vortex-type and U-plot drainage sink unit
11, the sample surface with UV light plane distance of 50 mm and parallel to the
Third, the heating system
1, U-shaped titanium alloy using high-speed electric heating tube
2, completely independent systems, does not affect the UV test and control lines
3, the temperature control output calculation by the microcomputer to achieve high precision and high efficiency power efficiency
4, with a heating system, anti-temperature functions
Fourth, the board temperature control
1, with a black aluminum plate temperature sensor connection
2, using the blackboard instrument control heating temperature, the temperature more stable
Fifth, the circuit control system
1, LCD digital display controllers, control accuracy ± 0.1 ℃
2, membrane KEY BOARD
3, with power and memory functions, when the restoration of the testing process can continue with the remaining
4, temperature control are used P. I. D + S. S. R, coordination and control system with the channel, and interface control elements can be used to improve the stability and life
6, with P. I. D automatic calculation of the function, can reduce the inconvenience caused when the artificial
7, light and condensation, spray can independently control the alternating cycle of control
8, independent control of light and condensation cycle time and turn time can be controlled by any of a thousand hours to set
9, in operation or settings, such as when an error occurs, it will provide rapid warning No.
10, years blackout recording device
11, France "Schneider" components.
12, Philips rectifier and non-starter (start time to ensure that UV light can be lit)
Six light
1, light source 8 is rated at 40W Philips fluorescent UV light source for hair. UV fluorescent tubes, located on both sides of the machine on each side 4. With UVA-340 and UVB-313 light source for the user to select the configuration. (Choose one)
2, UVA-340 lamp emission spectrum energy is concentrated in the wavelength 340nm
3, UVB-313 lamp emission spectrum mainly in the vicinity of 313nm wavelength.
4, the fluorescent light energy output will gradually decay over time, in order to reduce the attenuation of light caused by the impact test, so the chamber in all eight lamps in every 1 / 4 of the fluorescent lifetime, by a new branch replaces an old lamp tube, so that UV light source is always new and the old lamp from the composition, resulting in a constant output of light energy.
5, lamp life can be effective at 500 hours
Seven spray control
1, spray uniformity of regulation: the use of the controller manual control functions, observed in the open state spray performance, and thus to adjust or replace the nozzle.
2, spray condition monitoring: a spray device configured machine, spray device simulation of changes in temperature when it rains and water erosion, a total number of nozzles, spray evenly. When spraying, spray long, set free by the customer.
3, the user pressure to be prepared in deionized water.
Eight security guards
1, protective lock: If the lamp in the light, the cabinet door, once opened, the machine will automatically cut off the lamp power supply, cooling and automatically enter a state of equilibrium, so as not to harm the human body, safety locks meet IEC 047-5 - 1 security requirements.
2, the temperature inside the over-temperature protection: When the inside temperature exceeds 93 ℃ ± 10%, the machine will automatically cut off the tube and the heater power supply, and enter a state of equilibrium cooling.
3, tank low water level alarm to prevent air heater burning.
(The above is the leakage, short circuit, over-temperature, water shortages, over-current protection / memory controller power failure)
IX standard
GB/T14522-93 GB/T16422.3-1997 GB/T16585-96; UV aging tests such as the current standard.
X. equipment environment
1, Power: 220V/50HZ
2, Environment: 20 ~ 30 ℃ (ASTM G53 recommended), 0 ~ 80% RH, good ventilation, the indoor environment clean.
3, the working area: 234 × 353cm (including staff work space)
4, water: deionized water pressure 193Kpa around, 18.5 ℃, flow of about 7 l / min, requiring total insoluble content <10ppm.
5, the drain: the host is required on the floor near the drain.